SOP 7. Humidity Measurements
HMW61Y
Humidity Transmitter
Electrical Cable Connection

The probe of the transmitter should always point downwards when installed.
Supply voltage: 10 … 35 VDC (RL = 0 W)
20 … 35 VDC (RL = 500 W)
RH range: 0 …100% (Output 4…20mA)
T range: -20°C … +80°C (Output 4…20mA)
|
Temperature (°C) |
-20 |
-10 |
0 |
10 |
20 |
30 |
40 |
50 |
60 |
70 |
80 |
|
Output (mA) |
4 |
5.6 |
7.2 |
8.8 |
10.4 |
12 |
13.6 |
15.2 |
16.8 |
18.4 |
20 |
Reference
HMW61U/Y Humidity Transmitter
Operating Manual, March 2000.
HMK15
Humidity Calibrator
Introduction
The functioning of the HMK15 is based on the fact that certain salt
solutions generate a certain relative humidity in the air above them. The salt
solutions suitable for the HMK15 calibrator are e.g. lithium chloride LiCl (11%
RH), magnesium chloride MgCl2 (33% RH), sodium chloride NaCl (75%
RH) and potassium sulphate K2SO4 (97% RH). For
calibration, the sensor head is inserted into a salt chamber containing a
saturated salt solution. The probe/transmitter reading is then adjusted to the
correct value. Calibration is usually performed at least at two different
humidities to ensure the sensor accuracy over the entire humidity range (0-100
% RH).
|
Salt Name |
RH |
Usage |
|
LiCl |
11% |
Used as the dry end reference |
|
MgCl2 |
33% |
Used as a check point if calibration is performed at more
than two points. |
|
NaCl |
75% |
Used as the wet end reference for probes measuring in
applications with normal humidities. |
|
K2SO4 |
97% |
Used as the wet end reference for probes measuring in
applications with very high humidities. |
NOTE: Never add
water to dry LiCl salt.
LiCl is harmful when swallowed; the solution is also
corrosive.
If the LiCl solution is used or stored in temperatures
below +18 C, its equilibrium humidity changes permanently.
1. Take the calibrator out of the box. Open the transit
cover of the chamber. Remove the measurement cover from the chamber holder and
press the transit cover on the holder.
2. Pour ion exchanged water into the chamber; the
required amounts are given below:
|
LiCl |
14 ml of water |
|
MgCl2 |
3 ml of water |
|
NaCl |
10 ml of water |
|
K2SO4 |
10 ml of water |
3. Sprinkle the contents of a salt package in small
quantities into the chamber, stirring constantly. When measuring with the
measurement cup, make sure that the cup is clean and dry. Rinse and dry the cup
after every use.
|
LiCl |
15 g or 18 ml |
|
MgCl2 |
30 g or 30 ml |
|
NaCl |
20 g or 15 ml |
|
K2SO4 |
30 g or 20 ml |
4. When all salt has been sprinkled into the chamber, the
saturated salt solution should have the ratio of 60-90% undissolved salt to
10-40% liquid.
5. Close the chamber with the chamber cover. Fasten the
salt chamber on the base plate and close the measurement holes with rubber
plugs. Make sure that chamber covers
and plugs are carefully closed.
6. Write the preparation date on a sticker and mark the
chamber with it.
7. Allow approximately 24 hours for stabilization before
use.
Notices for Avoiding Errors Introduced
by Temperature Difference
Usually, the errors during humidity
calibration are due to temperature differences. In laboratory, the calibrator
should be stored in that part of the room where the temperature at most stable
and must be kept out of direct sunlight and away from localized heat sources.
Handle the probe as little as possible. Do
not hold the salt chamber or other parts of the calibrator in your hand during
calibration as they warm up and cause errors in the readings.
During calibration, the thermometer is
inserted into the 13.5 mm hole of a salt chamber. Press it downwards until it
passes the O-rings. The thermometer is correctly in place when you can feel a
resistance while pressing it downwards.
When the thermometers are not in use or the
calibrator is transformed from one place to another, place the thermometer in
holders.
Calibration procedure
(If the probe/transmitter is checked against
several humidity references, the checking must first be made at the dry end.)
Greenspan’s
calibration table
|
°C |
LiCl |
MgCl2 |
NaCl |
K2SO4 |
|
0 |
* |
33.7±0.3 |
75.5±0.3 |
98.8±1.1 |
|
5 |
* |
33.6±0.3 |
75.7±0.3 |
98.5±0.9 |
|
10 |
* |
33.5±0.2 |
75.7±0.2 |
98.2±0.8 |
|
15 |
* |
33.3±0.2 |
75.6±0.2 |
97.9±0.6 |
|
20 |
11.3±0.3 |
33.1±0.2 |
75.5±0.1 |
97.6±0.5 |
|
25 |
11.3±0.3 |
32.8±0.2 |
75.3±0.1 |
97.3±0.5 |
|
30 |
11.3±0.2 |
32.4±0.1 |
75.1±0.1 |
97.0±0.4 |
|
35 |
11.3±0.2 |
32.1±0.1 |
74.9±0.1 |
96.7±0.4 |
|
40 |
11.2±0.2 |
31.6±0.1 |
74.7±0.1 |
96.4±0.4 |
|
45 |
11.2±0.2 |
31.1±0.1 |
74.5±0.2 |
96.1±0.4 |
|
50 |
11.1±0.2 |
30.5±0.1 |
74.4±0.2 |
95.8±0.5 |
Quality control
Depending on the frequency of use and the
general operating conditions, the salt solutions should be replaced after about
6-12 months. A visual check should be performed at intervals of 2-3 months.
There must be a minimum of approximately 10% of undissolved salt at the bottom
of the chamber and the salt must be clean.
Note: LiCl may crystallize on the surface. But it
may still have solution under the surface. In such case, stir the solution and
check the next day.
For correct calibration, it is essential that
salt chambers are tightly closed. Check the O-rings at each salt replacement.
If they are damaged, replace with new ones.
Operating temperature range is 0-50°C.
Accuracy of salt solution humidities
Lithium chloride LiCl ±1.0%RH
+ Greenspan’s uncertainty
Magnesium
chloride MgCl2 ±1.0%RH
+ Greenspan’s uncertainty
Sodium chloride NaCl ±1.4%RH
+ Greenspan’s uncertainty
Potassium
sulphate K2SO4 ±1.5%RH
+ Greenspan’s uncertainty
Accuracy of the thermometer
With
mercury ±0.3°C
With
red capillary liquid ±1°C
Reference
HMK15
Humidity Calibrator Operating Manual, June 1998.
HMI41
Indicator and HMP46 Probe
Calibration Procedure
The calibration is
not needed unless the adjustment has been changed. The probe must always be
recalibrated when the sensor is changed.
A. Probe
calibration with the trimmer potentiometer:
The potentiometer marked “T” is for factory use only. “D” is for the dry end,
and “W” is for the wet end. Calibration can be performed with the HMK15.
1.
Leave the HMK15 calibrator and the probe at
the calibration site for at least 30 minutes to stabilize before starting the
calibration.
2.
Insert the probe into the measure hole of the
LiCl solution. Wait until the humidity reading stabilizes, this will take about
10-30 minutes. Then use a suitable wrench to adjust the “D” of the trimmer
potentiometer to the humidity value on the Greenspan’s calibration table.
3.
Then use the NaCl or K2SO4 as
the wet reference depends on the humidity we need to use. If calibrating probes
that are being used for a long time in high humidities (90-100RH), use the K2SO4.
Otherwise, use the NaCl. Adjust the “W” of the trimmer potentiometer to the
humidity value on the Greenspan’s calibration table. (Note that in high
humidities the risk for errors increases. Therefore the stabilization time
should be longer (about 20-40 min). )
B. Calibration with HMI41 Software
Commands:
This method is useful only when one probe is used,
otherwise, we should use the probe potentiometers.
Refer to page 20 of the manual.
Refer to page 26 of the manual.
Refer to page 33 of the manual.
Reference
HMI41 Indicator and HMP41/45/46 Probes Operating
Manual, March 1998.
Table 1. Maintenance log
|
Data |
Operator |
Reason (salt
replacement or other maintenance) |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Table 2. Salt solution checks
|
Data |
Operator |
LiCl |
MgCl2 |
NaCl |
K2SO4 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|