SOP 6. Real-time PM10 Monitor
The TEOM instrument (TEOM
1400a Ambient Particulate (PM-10) Monitor) is a continuous PM monitoring device
designated by USEPA as an equivalent method (EPA Designation No. EQPM-1090-079) for PM10 (10 microns sized particles and
under) and used extensively in state and national PM2.5 monitoring
networks. The acronym TEOM stands for “Tapered Element Oscillating
Microbalance,” an inertial measurement technique that operates on changes in
the resonant frequency of an oscillating element as a function of increases in
particle mass collected on a filter attached to the element. Changes in the
element’s resonant frequency are sampled electronically in quasi-real time,
providing both continuous and time-averaged measures of mass accumulation that
are directly proportional to instantaneous and time-averaged mass
concentrations in air, respectively. The device operates at an
industry-standard, volume-controlled flow rate of 16.7 L/min so that it can be
outfitted with a variety of commercially available pre-separator inlets
suitable for measuring PM10, PM2.5 or any other size
fraction of interest.
Since the airflow is
maintained at a constant volume, corrected for local temperature and barometric
pressure, the operation of this monitor requires that the temperature and
pressure sensors are connected for proper temperature and pressure readings and
flow corrections.
Supply power to the
instrument by plugging in the power cord to 120 VAC, and pressing the power
button on the front panel of the TEOM control unit. The main screen (four-line
display) will soon appear after showing the name of the instrument. Turn on the
pump to draw the sample streams by plugging in the power cord to 120 VAC. The
monitor waits at least 30 min after being powered up to compute the first mass
concentration data.
Press the power button on the
front panel of the TEOM control unit. The four-line display becomes blank. Turn
off the vacuum pump by pressing the power button and disconnect the control
unit from 120 VAC by unplugging the power cord.
Whenever a status code other
then “OK” is shown on the display, the instrument automatically turns on the
“Check Status” light on the front of the control unit. The information
displayed on the main screen includes status condition, operating mode, A/O 1
mode, RS-232 mode, protection, and time. For example, the screen will show the
following line at 8:11 AM:
OK 4+ 51% NU 08:11
Status condition:
OK Normal operation
M No frequency signal
T Temperature(s) outside of operational
bounds
F Flow(s) outside of operational bounds
X
Filter nearing capacity—exchange
filter
Operating mode:
1
Temperature/flow
stabilization
2
Begin TM
computation
3
TM computed,
begin MR/MC computation
4
Normal operating
mode
S Set up
mode
X Stop
all mode
A/O 1 mode:
(Blank) Analog output 1 normal definition
+ Analog output 1 used for status watch
XX% Filter
loading (percent)
RS-232 mode:
N None
P Print online
R R&P protocol
A AK protocol
G German network protocol
S Storage to printer
F Fast storage output
Protection:
U Unlocked
L Low lock
H High lock
XX:XX Current
time (24-hr format)
User should refer to Section
4.5 of the operating manual and become familiar with the incorporated
menu-driven software and keypads.
The measurements must be
conducted with TEOM filter cartridges that are made of Teflon-coated glass
fiber filter paper. Filters should be stored inside the sensor unit for easy
access and to keep them dry and warm. Do not handle new filter with your
fingers. Instead, use the filter exchange tool and follow the procedures
given in the operating manual. Keep the sample pump running to facilitate
filter exchange. Use the two pockets on the right side of the mass transducer
(inside) to store the next two new TEOM filters, for pre-conditioning and
removal of excessive moisture build-up prior to use.
The system resets itself when
power is regained, and enters the same RS-232 mode as before. All operating
parameters are maintained in the system’s battery backed-up CMOS memory.
Values logged internally are
stored in a circular buffer, and can be viewed from the control unit (press
<Store>, and select “View Storage”, or enter 08<Enter> from any
screen), or downloaded over the monitor’s RS-232 port. The instrument always
stores the time, date, and station number in each record in addition to the
data fields selected by the user. The storage capacity is given in Fig. 4-11 of
the operating manual, generally about 1.7 weeks of 8 data fields per record,
using a storage interval of 10 minutes.
The <Step Screen> key
toggles the instrument between the View Storage Screen and Set Storage Screen,
or, pressing 09<Enter> can gain direct access to the Set Storage Screen
from any screen. The first eight lines of the Set Storage Screen contain the
titles of the variables currently being stored in the data logger (Program
Register Codes listed in Appendix A of operating manual). The “Interval”
variable defines the time (in seconds) between successive writings of data to
the circular buffer.
Some popular program
register codes:
|
Code |
Variable |
Units |
|
008 |
Mass concentration |
mg/m3 |
|
009 |
Total mass |
mg |
|
035 |
Filter loading |
% |
|
039 |
Current main flow |
L/min |
|
040 |
Current auxiliary flow |
L/min |
|
041 |
Status condition |
code |
|
057 |
30-min average mass
concentration |
mg/m3 |
|
130 |
Current ambient temperature |
oC |
|
131 |
Current ambient pressure |
atm |
The instrument’s three analog
output channels are accessible from the identical 15-pin connectors on the
front and back panels of the control unit. Details regarding the pin
assignments and voltage (VDC) of outputs are given in the operating manual
(Section 5.1). To bring up the Set Analog Output Screen, press <A/O>, or
select “Set Analog Output” from Menu Screen, or press 04<Enter>.
Connect an IBM AT-compatible
computer to an RS-232 port using the provided 9-to-9 pin computer cable. Make
sure the computer is logged to the proper directory and subdirectory for
program execution. Type TEOMCOMM <Enter> to bring up the software. Use
the “Download storage” command to download the stored data in the data logger.
Ensure that the communication software is set for the same communication
parameters as the instrument.
Details of serial outputs and
two-way communications (such as online printing and connecting to a computer
through a modem) are given in section 6 of the operating manual.
TEOM Tapered
element oscillating microbalance
TM Total
mass
MR Mass
rate
MC Mass
concentration
AK AK
protocol was developed by the German automobile industry to standardize
thecommunication among instrumentation. This protocol is used in combination
with the TEOMCOMM software for data logging of the system.
A/O Analog
output
atm atmosphere,
746 mBar
Procedures are based on
routine flow auditing, leak checking, and mass calibration verification. Since
the TEOM monitor can be directly mass calibrated, it can be directly quality
assured using a mass standard. All QA procedures should be coordinated with
routine maintenance procedures to minimize down time.
A flow audit adapter is
provided and the procedures are outlined in the operating manual. Both the
sample flow rate and total flow rate may be checked using the flow audit
adapter with a capped nut for closing the flow splitter bypass line port. It is
recommended that the volumetric flow rates be within ±7% of the set points. The
United States Environmental Protection Agency (USEPA) requires a tolerance of
±10% for the total flow through the PM-10 inlet. If measured flows differ by more
than the stated tolerances, recheck all settings, and perform the test again.
Large errors in the flow may indicate other sources of error, such as a
malfunctioning flow controller, a system leak, or improper temperature and
pressure settings.
The leak check procedures are included in the operating manual (Section
7.6). The leak check should be performed with NO sample filter attached to the
mass transducer, which will prevent accidental damage from occurring to the
sample filter cartridge when exposed to the high pressure drop (vacuum) in the
sample line that the leak check creates. Flow rates should indicate less than
0.15 L/min for the main flow and less than 0.65 L/min for the auxiliary flow with
the end of the sample line closed, if not, systematically check plumbing for
connector leaks.
The mass transducer is permanently calibrated and never requires
recalibration under normal use. However, the mass measurement accuracy of the
instrument may be verified following procedures in the operating manual.
R&P offers a mass calibration verification kit to help perform this
procedure.
The lifetime of a TEOM filter cartridge depends on the nature
and concentration of the particulate sampled, and the main flow rate setting
(1, 2, or 3 l/min). The filter must be exchanged when the filter loading value
(as shown on the status line of the main menu) approaches 100%. At a
flow rate of 3 l/min, 100% filter loading generally corresponds to a total mass
accumulation of approximately 3 to 5 mg of particulates. Filter lifetime at a
main flow rate of 3 l/min is generally 21 days at an average PM-10
concentration of 50 mg/m3. Flow splitter adapters for 1 and 2
l/min operation are available for use in areas with higher particulate
concentrations.
The factory recommended schedule of periodic maintenance and the
schedule to be adopted for use in livestock buildings are as follows:
|
Maintenance item |
Factory |
APECAB |
|
Clean external screen |
- |
As necessary |
|
Clean PM-10 inlet |
Upon filter
exchange |
Weekly |
|
Exchange in-line filters |
6 mon. or when
loaded |
When loaded |
|
Clean air inlet system |
6 mon. |
Monthly |
|
Leak test |
Annually |
Bimonthly |
|
Mass flow controller calibration |
Annually |
Bimonthly |
|
Analog board calibration |
Annually |
Every 5 months |
|
Mass calibration verification |
Annually |
Bimonthly |
The PM10 inlet requires regular maintenance in livestock buildings. A
wire-mesh external screen was designed for the PM10 inlet to keep large dust
particles out. This external screen requires cleaning every two to three days,
depending on PM concentration in the air. Cleaning is accomplished by carefully
removing the external screen, and, in an area downwind of the inlet unit,
brushing it or cleaning with water and letting dry. After all visible dust has
been removed from the external screen, carefully replace it on the PM10 head,
so as not to disturb the sensor unit.

Figure. TEOM PM10 inlets with external screens. (Source: Purdue
University)
The PM10 inlet itself should be cleaned weekly. In order to
do this one must first push the <Data Stop> button on the keypad. Remove
the PM10 inlet and replace it with an in-line filter to prevent excessive
amounts of dust being pulled into the TEOM unit. At this point, unscrew the
Acceleration Assembly from the Collector Assembly (see fig. F-1). Using a can
of pressurized Dust Remover (found at most hardware and electronics retailers),
spray the inlet screen on the Acceleration Assembly until clean. Then spray the
inside of the Acceleration Assembly from the bottom tube. This should remove
most of the dust from the screen. Also spray the internal areas of the
Collector Assembly. After these have been cleaned, feel the o-rings, located on
the lower part of the Acceleration Assembly and on the lower part of the
Collector Assembly, for the presence of silicon grease. Grease as needed to
maintain a slight layer of Silicon Grease. Carefully replace the inlet unit,
and return to the control unit of the TEOM system, and press the <Data
Stop> button once more. This will prevent skewed data from being recorded. It
will automatically begin recording in approximately 30 min.
Flow rate through sample
inlet 16.7 L/min (1 m3/hr)
Main flow rate 3 L/min
Temperature of sample stream 50 oC
Particulate concentration <
5 mg/m3 to several g/m3
Slow blow 2A, 250 V fuse
2A $ 250 V in-line fuse, P/N
04003419
Box of 20 TEOM filter
cartridges (TX40 media), P/N 57-000397-0020
Large bypass in-line filter,
P/N 57-002758
Flow controller filter, P/N
30-003097
Manufacturer
Contact Information
Rupprecht &
Patashnick Co., Inc.
25 Corporate Circle
Timothy Morphy
Assistant Product Manager
518.452.0065 voice
518.452.0067 fax
tmorphy@rpco.com (email)
Peggy O’Gormon
Marketing Associate
X3229, email:
pogorman@rpco2.com