SOP 4
Pulsed-fluorescence hydrogen sulfide analyzer
(TEI Model 45C)
Hydrogen sulfide is converted catalytically
at (200-400 °C) to SO2 with a converter (TEI Model 340) followed by measurement with
a pulsed fluorescence SO2 detector (TEI Model 45C) (U.S. EPA. Method
EQSA-0486-060) equipped with a high intensity xenon lamp. The analyzer is based
on the principle that SO2 molecules absorb ultraviolet (UV) light
and becomes excited at one wavelength, decay to a lower energy state, and emit
UV light at a different wavelength (Equation 2).
SO2 + hv1 ® SO2* ® SO2 + hv2 (2)
where * refers to a
molecule in an excited state. The emitted UV light is proportional to SO2
concentration and is detected by a PMT. The SO2 analyzer has a range
of 0.05 to 100 ppm, a response time of 60 s with a 10 s averaging time, and a
sample flow rate of 0.5 L/min. Full scale is
adjustable up to 100 ppm. The data averaging time is adjustable up to 60
s. The guaranteed precision is 1% of reading or 1 ppb (whichever is greater)
and the linearity is ±1% of full scale (TEI Model 340 H2S Converter Instruction Manual and Model 43C Pulsed Fluorescence SO2
Analyzer Instruction Manual).
Zero air, sulfur dioxide (SO2) in N2 and H2S
[JN1]in
N2 are used to calibrate the instrument. Unless SO2 is one of the
target analytes, it is only used to check converter efficiency.
a. Select calibration
press ENTER
b. Select calibrate
zero press ENTER
c. When readings are stable press ENTER to zero the analyzer
d. Press RUN
1) Close regulator valve on span gas cylinder
2) Open main valve on the SO2 cylinder
3) Insert 1/8” ID tubing (from gas cylinder) into the
calibration manifold connected to the valve; then open regulator valve to allow
gas flow
4) Adjust regulator valve until gas flow is 1 L/min (read
from bottom of ball of the flow meter).
5) Record the time and analyzer display on the lab notebook
after display is stabilized (typically 5 to 10 minutes).
6) Press MENU
on the SO2 Analyzer
a. Select calibration
press ENTER
b. Select SO2
press ENTER
c. Set SO2 value to concentration of the
calibration gas
d. When readings are stable press ENTER
e. Press RUN
7) Close regulator and remove tubing from gas cylinder
1) Close regulator valve on the span gas cylinder
2) Open main valve on span gas cylinder
3) Insert the 1/8” ID tubing (from gas cylinder) into the
calibration manifold then open the regulator valve to allow gas flow
4) Adjust regulator valve until airflow is 1 L/min (read
from bottom of ball of the flow meter).
5) Wait for 5 to 10 minutes for the display to stabilize,
record the time and analyzer display on the lab notebook
6) Press MENU
on the H2S Analyzer
a. Select calibration
press ENTER
b. Select H2S press ENTER
c. Set H2S value to the concentration of the
calibration gas
d. When readings are stable press ENTER
e. Press RUN
7) Close regulator and remove tubing from gas cylinder
Reattach the filters to the existing gas sampling
system.
Check and close main valves on all cylinders
Calibration of the H2S analyzer
is complete
List of Spare
Parts (to be completed)
Manufacturer
Contact Information
Thermo Environmental Instruments
(www.thermoei.com)
Tel: (508) 520-0430 ext. 6812
Fax: (508) 520-1460
Sales: Souphin A. Sithideth, Inside Sales
Engineer, e6812, E-Mail: SSithideth@thermoei.com
Service: Barry Pepin, Service Engineer, e6908, E-Mail:
Bpepin@thermoei.com
|
Time |
Items |
Unit = |
Notes |
|
|
: : |
Sample airflow (L/min) |
|
||
|
: : |
Zero air applied |
SO2 |
H2S |
Cylinder P: _____ psi |
|
: : |
|
|
|
|
|
: : |
|
|
|
|
|
: : |
Reset to zero gas |
Yes/No |
|
|
|
: : |
SO2 ( __ ppm) applied |
--- |
--- |
Cylinder P: _____ psi |
|
: : |
|
|
|
|
|
: : |
|
|
|
|
|
: : |
Reset to span gas |
Yes/No |
||
|
: : |
H2S ( __ ppm) applied |
--- |
--- |
Cylinder P: _____ psi |
|
: : |
|
|
|
|
|
: : |
|
|
|
|
|
: : |
Reset to span gas |
Yes/No |
||
|
: : |
New coefficient saved |
Yes/No |
||
|
: : |
Connect analyzers back to the CEM system. |
|||
[JN1]
NH3 measurement will be set at 0-50 ppm.